The YE0175 is designed for measurement of the disturbance caused by the video level in black areas to the right of or below the areas. Horizontal white lines are positioned on a black background.
Definition of streaking measurements
Measurement of the disturbance caused by the video level in black areas to the right of or below the areas.The disturbance is characterized by measuring the duration of the delay before the video level corresponding to the normal video level of the black is restored.
Measuring equipment
Video level meter; B/W monitor, oscilloscope, low-pass filter: 500 KHz
Measurement conditions
The chart illumination must be adjusted so that the white bars give a signal of 100% with average iris setting.
Gain: 0dB
Gamma corrector: OFF
Contour corrector: OFF
Knee corrector: OFF
Pedestal = 5% of nominal video level (= 35 mV)
With 12 rigorous quality inspection processes in place, we ensure a high yield rate of up to 98% for batch processed precision parts.
FAQs About Other Charts
Always a pre-production sample before mass production; Always final Inspection before shipment;
Yes. We can supply OEM and make customized design for any specific application. Our MOQ for OEM and ODM service is 5 pcs for one model. We can not only offer standard machines, but also customized machines. Don’t hesitate to tell us your requirement, we will try our best to meet your needs.
a. If this phenomenon occurred, it means the battery power is not enough. You need to charge the battery.
b. When charge the battery, please insert the battery to the instrument, and plug in the adapter to charge the battery. After 5 minutes, the instrument can be turned on.
c. If the instrument still cannot be turned on after charging 5 minutes, you can pull out the adapter and re-plug in to check it. Or repeat this operation.
d. If the instrument still cannot be turned on after charging 5 minutes, you can turn off the colorimeter and turn on it again to check it. Or repeat this operation.
e. If you have tried all the methods above but the colorimeter still cannot be turned on, please change the battery.
a. Before installation, please Copy CQCS3 folder to your computer and don’t remove any files in CQCS3 folder.
b. Double-click the setup.exe to install CQCS3 software.
c. Then double-click Install.bat under the CQCS3USB_Driver file to install the driver.
d. Detailed installation information please read the "Installing Software" under CQCS3 User manual.doc.
a. Check whether the USB cable is connected. If not, please connect it. Check whether the connection is effective. You can re-plug the USB cable to test.
b. If the USB cable is connected well but the interface still displays "USB Disconnect", please re-start the colorimeter and enter "Menu" -"Comm"to start the communication.
c. Change the USB cable or change the USB port in you PC.